Human RDH11 (Retinol dehydrogenase 11) ELISA Kit

Cat# EH11754

Size : 96T

Brand : FineTest

Request more information


This kit is based on Double antibody-Sandwich ELISA detection method and takes 4h assay time. The microplate provided in this kit has been precoated with anti RDH11 antibody. Add standard and properly diluted sample into relevant well respectively. After incubation, wash unbound components. Add biotinylated detection antibody. Then, it binds with RDH11 bound to precoated antibody. Wash unbound components and add HRP-Streptavidin Conjugate (SABC). Wash unbound components again and add TMB substrate solution. Then, TMB was catalyzed by HRP to produce a blue color product that turned yellow after adding a stop solution. Read the O.D. absorbance at 450nm in a microplate reader. Calculate the concentration of RDH11 in the sample by plotting standard curve. The concentration of the target substance is proportional to the OD450 value.

Catalogue No.: 
EH11754

Alias: 
Retinol dehydrogenase 11 ELISA Kit, Androgen-regulated short-chain dehydrogenase/reductase 1 ELISA Kit, HCV core-binding protein HCBP12 ELISA Kit, Prostate short-chain dehydrogenase/reductase 1 ELISA Kit, Retinal reductase 1 ELISA Kit, RalR1 ELISA Kit, Short chain dehydrogenase/reductase family 7C member 1 ELISA Kit, RDH11 ELISA Kit, ARSDR1 ELISA Kit, PSDR1 ELISA Kit, SDR7C1 ELISA Kit

Species: 
Human

Range: 
78.125-5000pg/ml

Sensitivity: 
46.875pg/ml

  • SPECIFICATIONS
  • FIGURES
  • CONDITIONS
Product Name
Human RDH11 (Retinol dehydrogenase 11)ELISA Kit

Alias
Retinol dehydrogenase 11 ELISA Kit, Androgen-regulated short-chain dehydrogenase/reductase 1 ELISA Kit, HCV core-binding protein HCBP12 ELISA Kit, Prostate short-chain dehydrogenase/reductase 1 ELISA Kit, Retinal reductase 1 ELISA Kit, RalR1 ELISA Kit, Short chain dehydrogenase/reductase family 7C member 1 ELISA Kit, RDH11 ELISA Kit, ARSDR1 ELISA Kit, PSDR1 ELISA Kit, SDR7C1 ELISA Kit

Catalogue No.
EH11754

Size
48T/96T

Species
Human

UniProt ID
Q8TC12

Sample Type
Serum, Plasma, Cell Culture Supernatant, cell or tissue lysate, Other liquid samples

Detection Method
Sandwich ELISA, Double Antibody

Detection Wavelength
OD450

Reaction Duration
4 hours

Range
78.125-5000pg/ml

Sensitivity
46.875pg/ml

Storage
2-8°C(Sealed), Don't cryopreserve.

Specificity
Specifically binds with RDH11 , no obvious cross reaction with other analogues.

ELISA Kit Components
Kit Components Item Size(48T) Size(96T) Storage Condition for Opened Kit
E001 ELISA Microplate(Dismountable) 8×6 8×12 Put the rest strips into a sealed foil bag with the desiccant. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E002 Lyophilized Standard 1vial 2vial Put the rest standards into a desiccant bag. Stored for 1 month at 2-8°C; Stored for 6 month at -20°C
E003 Biotin-labeled Antibody(Concentrated, 100X) 60ul 120ul 2-8°C (Avoid Direct Light)
E034 HRP-Streptavidin Conjugate(SABC, 100X) 60ul 120ul
E024 TMB Substrate 5ml 10ml
E039 Sample Dilution Buffer 10ml 20ml 2-8°C
E040 Antibody Dilution Buffer 5ml 10ml
E049 SABC Dilution Buffer 5ml 10ml
E026 Stop Solution 5ml 10ml
E038 Wash Buffer(Concentrated, 25X) 15ml 30ml
E006 Plate Sealer 3 pieces 5 pieces  
E007 Product Description 1 copy 1 copy

Required Instruments and Reagents
  1. Microplate reader (wavelength: 450nm)
  2. 37°C incubator (CO2 incubator for cell culture is not recommenced.)
  3. Automated plate washer or multi-channel pipette/5ml pipettor (for manual washing purpose)
  4. Precision single (0.5-10μL, 5-50μL, 20-200μL, 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
  5. Sterile tubes and Eppendorf tubes with disposable tips
  6. Absorbent paper and loading slot
  7. Deionized or distilled water

Assay Procedure Summary
Elisa实验原理图
  • Step 1: Add 100ul standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37°C.
  • Washing: Wash the plate twice without immersing.
  • Step 2: Add 100ul biotin-antibody working solution, seal the plate and statically incubate for 60 minutes at 37°C.
  • Washing: Wash the plate three times and immerse for 1min each time.
  • Step 3: Add 100ul HRP-Streptavidin Conjugate (SABC) working solution, seal the plate and statically incubate for 30 minutes at 37°C.
  • Washing: Wash the plate five times and immerse for 1min each time.
  • Step 4: Add 90ul TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37°C. (Accurate TMB visualization control is required.)
  • Step 5: Add 50ul stop solution. Read at 450nm immediately and calculate.

Standard Curve

This product is detected by QC department and meets performance required in the manual. (Laboratory Humidity: 20%-60%; Temperature: 18°C -25°C; Equilibrate TMB substrate to 37°C before staining. After adding into the ELISA wells, incubate for 15min at 37°C in dark.)

Due to different assay environments and operations, assay data below and standard curve are provided for reference. Experimenters should establish standard curve according to their own assay.

STD.(pg/ml) OD-1 OD-2 Average
0 0.048 0.05 0.049
78.125 0.12 0.124 0.122
156.25 0.225 0.231 0.228
312.5 0.402 0.414 0.408
625 0.597 0.615 0.606
1250 0.912 0.938 0.925
2500 1.507 1.551 1.529
5000 2.432 2.502 2.467
EH11754 Standard Curve Image

Recovery

Add a certain amount of RDH11 into the sample. Calculate the recovery by comparing the measured value with the expected amount of RDH11 in the sample.

Sample Type Recovery Range(%) Average(%)
serum(n=10) 90-103 98
EDTA plasma(n=10) 86-104 94
Heparin plasma(n=10) 86-99 92

Linearity

Dilute the sample with a certain amount of RDH11 at 1:2, 1:4 and 1:8 to get the recovery range.

Sample Type 1:2 1:4 1:8
serum(n=10) 87-102% 86-95% 86-97%
EDTA plasma(n=10) 86-97% 87-99% 83-98%
Heparin plasma(n=10) 82-94% 80-97% 80-98%

Precision(%)

Intra-assay Precision: samples with low, medium and high concentration are tested 20 times on the same plate.

Inter-assay Precision: samples with low, medium and high concentration are tested 20 times on three different plates.

Item Intra-assay Precision Inter-assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 20 20 20
Mean (pg/ml) 158.1 599 2498 162.1 621 2530
Standard deviation 8.54 31.63 117.41 7.55 33.78 136.62
CV(%) 5.4 5.28 4.7 4.66 5.44 5.4

Stability

Perform the stability test for the sealed kit at 37°C and 2-8°C and get relevant data.

ELISA kit(n=5) 37°C for 1 month 2-8°C for 6 months
Average(%) 80 95-100